Technical standards on water quality have been established by a number of organizations including the American Society for Testing and Materials (ASTM), International Organization for Standardization (ISO), United States Pharmacopoeia (USP) and the Clinical and Laboratory Standards Institute (CLSI) - previously known as the US National Committee for Clinical Laboratory Standards (NCCLS). CLSI outlines Clinical Laboratory Reagent Water (CLRW) guidelines.
ASTM Standards for Laboratory Reagent Water (ASTM D1193-91)
ASTM: American Society for Testing and Materials
Measurement (Unit) | Type I | Type II | Type III | Type IV |
---|---|---|---|---|
Resistivity (MΩ-cm) | > 18 | > 1 | > 4 | > 0.2 (200KΩ) |
Conductivity (µS/cm) | < 0.056 | < 1 | < 0.25 | < 5 |
pH at 25˚C | N/A | N/A | N/A | 5.0 - 8.0 |
Total Organic Carbon (TOC) ppb or µg/L | < 50 | < 50 | < 200 | N/A |
Sodium (ppb or µg/L) | < 1 | < 5 | < 10 | < 50 |
Chloride (ppb or µg/L) | < 1 | < 5 | < 10 | < 50 |
Silica (ppb or µg/L) | < 3 | < 3 | < 500 | N/A |
The ASTM standards are further subdivided into A, B and C that can be used in conjunction with the type I, II, III or IV water above when bacteria levels need to be controlled.
Additional ASTM Sub-Standards for Laboratory Reagent Water
Measurement (Unit) | A | B | C |
---|---|---|---|
Heterotrophic Bacteria Count (CFU/ml) | < 1 | < 10 | < 1000 |
Endotoxin (units per ml) | < 0.03 | < 0.25 | N/A |
ISO 3696 Standard
ISO: International Organization for Standardization
Parameter | Grade 1 | Grade 2 | Grade 3 |
---|---|---|---|
Conductivity µS/cm (temp corrected) | < 0.1 | < 1 | < 5 |
pH at 25°C | N/A | N/A | 5.0 - 7.0 |
Oxidizable matter Oxygen (02) content mg/L | N/A | < 0.08 | < 0.4 |
Absorbance at 254 nm and 1 cm optical path length, absorbance units | < 0.001 | < 0.01 | N/A |
Residue after evaporation on heating at 110°C mg/kg | N/A | < 1 | < 2 |
Silica (Si02) mg/L | < 0.01 | < 0.02 | < N/A |
CLSI-CLRW Guidelines
CLSI-CLRW: Clinical and Laboratory Standards Institute - Clinical Laboratory Reagent Water
CLSI was formerly known as NCCLS (US National Committee for Clinical Laboratory Standards)
Contaminant | Parameter and Unit | Type 3 | Type 2 | Type 1 | CLRW |
---|---|---|---|---|---|
Ions | Resistivity (MΩ-cm) | > 0.05 (50 KΩ) | > 1 | > 18 | > 10 |
Organics | Total Organic Carbon (TOC) ppb | < 200 | < 50 | < 10 | < 500 |
Pyrogens | (Eu/ML) | N/A | N/A | < 0.03 | --- |
Particles | Particles > 0.2 µm (units/mL) | N/A | N/A | < 1 (0.22 µ filtration required) | Include 0.22 µ filtration |
Colloids | Silica (ppb) | < 1000 | < 100 | < 10 | --- |
Bacteria | Bacteria (cfu/mL) | < 1000 | < 100 | < 1 | < 10 |
These values are best used as guidelines, as many applications require further treatment based on other factors. For example, many molecular biology applications require Type 1 water that is free of DNase and RNase and simple washing of instruments (usually Type 3) might require water that is pyrogen free for critical applications (Type 1).
Laboratory Water Purity Specifications 'Consolidated' Guidelines
Contaminant | Parameter and Unit | Type 1 | Type 2 | Type 3 |
---|---|---|---|---|
Ions | Resistivity (MΩ-cm) | > 18 | > 1 | > 0.05 (50 KΩ) |
Silica (ppb) | < 10 | < 100 | < 1000 | |
Organics | Total Organic Carbon (TOC) ppb | < 20 | < 50 | < 200 |
Particles | Particles > 0.2 µm (#/ml) | < 1 | N/A | N/A |
Bacteria | Particles > 0.2 µm (#/ml) | < 1 | < 100 | < 1000 |
Endotoxin (Eu/ML) | < 0.001 | N/A | N/A |
Typical uses for each type are outlined below:
Type 1 | Required for critical laboratory applications such as HPLC, Mobile Phase Preparation, blanks and sample dilution for analytical techniques such as GC, HPLC, AA, ICP-MS, etc. Preparation of buffers and media for mammalian cell culture and IVF. Production of reagents for molecular biology applications (DNA sequencing, PCR). Preparation of solutions for electrophoresis and blotting. |
Type 2 | Used in buffers, pH solutions and microbiological culture media preparation and for preparation of reagents for chemical analysis. Used in clinical analyzers, cell culture incubators and weatherometers, etc. Also used as feed water to Type 1 systems. |
Type 3 | Used for glassware rinsing, filling autoclaves and heating baths and humidifiers. Also used as feed water to Type 1 systems. |
Source: www.millipore.com
USP Standards
USP: United States Pharmacopoeia
Properties | USP 'Purified Water' | USP 'Water for Injection' & 'Highly Purified Water' |
---|---|---|
Conductivity (µS/cm @ 25°C) | < 1.3 | < 1.3 |
Total Organic Carbon (TOC) ppb or µg/L | < 500* | < 500* |
Bacteria (guideline) | < 100 cfu/ml | < 10 cfu/100ml |
Endotoxin (EU/ml) | N/A | 0.25 EU/ml |
*Or pass oxidisable substance test