Laboratory Water Quality Standards

Technical standards on water quality have been established by a number of organizations including the American Society for Testing and Materials (ASTM), International Organization for Standardization (ISO), United States Pharmacopoeia (USP) and the Clinical and Laboratory Standards Institute (CLSI) - previously known as the US National Committee for Clinical Laboratory Standards (NCCLS). CLSI outlines Clinical Laboratory Reagent Water (CLRW) guidelines.

ASTM Standards for Laboratory Reagent Water (ASTM D1193-91)

ASTM: American Society for Testing and Materials

Measurement (Unit) Type I Type II Type III Type IV
Resistivity (MΩ-cm) > 18 > 1 > 4 > 0.2 (200KΩ)
Conductivity (µS/cm) < 0.056 < 1 < 0.25 < 5
pH at 25˚C N/A N/A N/A 5.0 - 8.0
Total Organic Carbon (TOC) ppb or µg/L < 50 < 50 < 200 N/A
Sodium (ppb or µg/L) < 1 < 5 < 10 < 50
Chloride (ppb or µg/L) < 1 < 5 < 10 < 50
Silica (ppb or µg/L) < 3 < 3 < 500 N/A

The ASTM standards are further subdivided into A, B and C that can be used in conjunction with the type I, II, III or IV water above when bacteria levels need to be controlled.

Additional ASTM Sub-Standards for Laboratory Reagent Water

Measurement (Unit) A B C
Heterotrophic Bacteria Count (CFU/ml) < 1 < 10 < 1000
Endotoxin (units per ml) < 0.03 < 0.25 N/A

ISO 3696 Standard

ISO: International Organization for Standardization

Parameter Grade 1 Grade 2 Grade 3
Conductivity µS/cm (temp corrected) < 0.1 < 1 < 5
pH at 25°C N/A N/A 5.0 - 7.0
Oxidizable matter Oxygen (02) content mg/L N/A < 0.08 < 0.4
Absorbance at 254 nm and 1 cm optical path length, absorbance units < 0.001 < 0.01 N/A
Residue after evaporation on heating at 110°C mg/kg N/A < 1 < 2
Silica (Si02) mg/L < 0.01 < 0.02 < N/A

CLSI-CLRW Guidelines

CLSI-CLRW: Clinical and Laboratory Standards Institute - Clinical Laboratory Reagent Water
CLSI was formerly known as NCCLS (US National Committee for Clinical Laboratory Standards)

Contaminant Parameter and Unit Type 3 Type 2 Type 1 CLRW
Ions Resistivity (MΩ-cm) > 0.05 (50 KΩ) > 1 > 18 > 10
Organics Total Organic Carbon (TOC) ppb < 200 < 50 < 10 < 500
Pyrogens (Eu/ML) N/A N/A < 0.03 ---
Particles Particles > 0.2 µm (units/mL) N/A N/A < 1 (0.22 µ filtration required) Include 0.22 µ filtration
Colloids Silica (ppb) < 1000 < 100 < 10 ---
Bacteria Bacteria (cfu/mL) < 1000 < 100 < 1 < 10

These values are best used as guidelines, as many applications require further treatment based on other factors. For example, many molecular biology applications require Type 1 water that is free of DNase and RNase and simple washing of instruments (usually Type 3) might require water that is pyrogen free for critical applications (Type 1).

Laboratory Water Purity Specifications 'Consolidated' Guidelines

Contaminant Parameter and Unit Type 1 Type 2 Type 3
Ions Resistivity (MΩ-cm) > 18 > 1 > 0.05 (50 KΩ)
  Silica (ppb) < 10 < 100 < 1000
Organics Total Organic Carbon (TOC) ppb < 20 < 50 < 200
Particles Particles > 0.2 µm (#/ml) < 1 N/A N/A
Bacteria Particles > 0.2 µm (#/ml) < 1 < 100 < 1000
  Endotoxin (Eu/ML) < 0.001 N/A N/A

Typical uses for each type are outlined below:

Type 1 Required for critical laboratory applications such as HPLC, Mobile Phase Preparation, blanks and sample dilution for analytical techniques such as GC, HPLC, AA, ICP-MS, etc. Preparation of buffers and media for mammalian cell culture and IVF. Production of reagents for molecular biology applications (DNA sequencing, PCR). Preparation of solutions for electrophoresis and blotting.
Type 2 Used in buffers, pH solutions and microbiological culture media preparation and for preparation of reagents for chemical analysis. Used in clinical analyzers, cell culture incubators and weatherometers, etc. Also used as feed water to Type 1 systems.
Type 3 Used for glassware rinsing, filling autoclaves and heating baths and humidifiers. Also used as feed water to Type 1 systems.

Source: www.millipore.com

USP Standards

USP: United States Pharmacopoeia

Properties USP 'Purified Water' USP 'Water for Injection' & 'Highly Purified Water'
Conductivity (µS/cm @ 25°C) < 1.3 < 1.3
Total Organic Carbon (TOC) ppb or µg/L < 500* < 500*
Bacteria (guideline) < 100 cfu/ml < 10 cfu/100ml
Endotoxin (EU/ml) N/A 0.25 EU/ml

*Or pass oxidisable substance test

Ready to Get Started? Contact Us Now or call 24/7 (800) 906-6060

Puretec provides high purity water solutions for industrial applications across the Southwestern United States.

Oxnard, California
3151 Sturgis Road
Oxnard, CA 93030

Serving Ventura, Santa Barbara, Los Angeles, and Kern Counties.

Anaheim, California
1815 S. Lewis St
Anaheim, CA 92805

Serving Orange, Los Angeles, San Bernadino, and Riverside Counties.

San Diego, California
7151 Ronson Road
San Diego, CA 92111

Serving San Diego, Imperial, and Riverside Counties.

San Jose, California
1291 Oakland Road
San Jose, CA 95112

Serving Santa Clara, San Francisco, San Mateo, and Sacramento Counties.

Gilbert, Arizona
1440 N. Hobson Street
Gilbert, AZ 85233

Serving Maricopa, Yuma, Gila, Pinal, and Pima Counties.